22.11.6.1 Measurement accuracy example
The following example illustrates the effect percent-of-reading and offset have on measurement
accuracy. The effect of offset is usually negligible on large signals.
Example:
l
Sensor-signal voltage: approximately 2500 mV
l
CRBasic measurement instruction: VoltDiff()
l
Programmed input-voltage range (Range) : mV 2 500 (±-100 to 2500 mV)
l
Input measurement reversal (RevDiff): True
l
Data logger circuitry temperature: 10° C
Accuracy of the measurement is calculated as follows:
accuracy = percent-of-reading + offset
where
percent-of-reading = 2500 mV • ±0.04%
=±1 mV
and
offset = ±20 µV
Therefore,
accuracy = ±(1 mV + 20 µV) = ±1.02 mV
22.11.7 Minimizing offset voltages
Voltage offset can be the source of significant error. For example, an offset of 3 μV on a 2500 mV
signal causes an error of only 0.00012%, but the same offset on a 0.25 mV signal causes an error
of 1.2%. Measurement offset voltages are unavoidable, but can be minimized. Offset voltages
originate with:
l
Ground currents. See Minimizing ground potential differences(p. 190).
l
Seebeck effect
l
Residual voltage from a previous measurement
22. Tips and troubleshooting199

Table of Contents

Other manuals for Campbell CR310

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Campbell CR310 and is the answer not in the manual?

Campbell CR310 Specifications

General IconGeneral
BrandCampbell
ModelCR310
CategoryData Loggers
LanguageEnglish

Related product manuals