SIRIUS
®
TECHNICAL REFERENCE MANUAL
5.7.1. General counter specifications
Counter
Timebase
102.4 MHz
Time base accuracy Typical
5 ppm, Max: 20 ppm
Max. Bandwidth
10 MHz
Input Filter
500 ns, 1 μs, 2 μs, 4 μs, 5 μs and 7.5 μs
Input Level Compatibility
TTL (Low: <0.8 V, High >2 V)
Input Impedance
100 pull-up to +3.3 V
Input Protection
±25 Volt continuous
Alarm output
Open collector, max. 100 mA / 30 Volt
Sensor supply
5 V / 100 mA; 12 V / 50 mA
Warning
Be careful with voltages >25 VAC or >35 VDC! These voltages are already high enough in order to
get a perilous electric shock by touching the wiring.
Caution
GND of the counter input is connected to the GND of the analog channel.
Caution
For ACC+, CHG+ channels GND of the counter input is connected via a 50 Ω resistor to In- of the
analog channel.
SIRIUS
®
V20-1 146 / 336

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT SIRIUS Series and is the answer not in the manual?

DEWESOFT SIRIUS Series Specifications

General IconGeneral
CategoryData Loggers
ManufacturerDEWESOFT
SeriesSIRIUS
Resolution24-bit
InterfaceUSB, EtherCAT
SoftwareDEWESoft X
ChannelsExpandable
Sampling RateUp to 200 kS/s per channel
Input TypesVoltage, Current, IEPE, Strain, Thermocouple, RTD, Resistance, Digital, Counter
Input Voltage Range±10 V, ±200 V
IsolationChannel-to-channel and channel-to-ground
ConnectivityEthernet, USB
Operating Temperature-10 °C to +50 °C
Storage Temperature-20 °C to +70 °C
Power Supply10 to 30 V DC
DimensionsVaries depending on module configuration
WeightVaries depending on module configuration
Analog InputsYes, depending on module configuration