EasyManua.ls Logo

DEWESOFT SIRIUS Series

DEWESOFT SIRIUS Series
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
/
SIRIUS
®
TECHNICAL REFERENCE MANUAL
5.22.3. HS-STG: Sensor connection
To minimize electromagnetic influence, it is recommended to use twisted pair cables. The shield of the
cable should be connected to GNDiso on pin 4 of the DSUB-9 connector.
HS-STG Sensor connection
HS-STG Sensor ¼ Bridge connection
Hint
When your application requires a bandwidth 100 kHz, it is recommended to use external
bridge completion. This fully symmetrical system configuration will guarantee the best
possible noise reduction.
SIRIUS
®
V20-1 251 / 336

Table of Contents