5.6. SIRIUS® Waterproof specifications
SIRIUSiwe-6xSTGM-2xSTGM+
SIRIUSwe-HD-16xSTGS
SIRIUSiwe-6xLV-2xLV+
Analog inputs
8 ch voltage
Full / half / quarter bridge strain
16 ch voltage
Full / half / quarter bridge strain
8 ch voltage
Full bridge strain, current (ext.
Shunt)
ADC type
2x 24-bit delta-sigma
DUALCOREADC®
with anti-aliasing filter
24-bit delta-sigma
with anti-aliasing filter
2x 24-bit delta-sigma
DUALCOREADC®
with anti-aliasing filter
Sampling rate on USB
Simultaneous 200 kS/sec
Simultaneous 200 kS/sec
Simultaneous 200 kS/sec
Sampling rate on EtherCA
Simultaneous 20 kS/sec
Simultaneous 10 kS/sec
Simultaneous 20 kS/sec
Ranges
±10 V, ±1 V, ±100 mV, ±10 mV
±10 V, ±1 V, ±100 mV, ±10 mV
±200 V, ±10 V, ±1 V, ±100 mV
Bridge ranges
2 mV/V ... 1000 mV/V, free
programmable
2 mV/V ... 1000 mV/V, free
programmable
2 mV/V ... 1000 mV/V, free
programmable
Dynamic Range@10kS
-152 dB
-137 dB
-152 dB
Input coupling
DC
DC
DC, AC 1 Hz (3 Hz, 10 Hz per SW)
Input impedance
10
10
200 V Range: 1 MΩ;
all other ranges 10
between IN+ or In- against
GND
Bridge modes
Full / Half / Quarter Br 120 / 350
Ω 3-wire; internal bridge
completion
Full / Half / Quarter Br 120 / 350
Ω 3-wire; internal bridge
completion
Full
Internal shunt resistor
100 kΩ, bipolar to Exc+ or Exc-
(others on request)
100 kΩ, bipolar to Exc+ or Exc-
(others on request)
-
Excitation voltage
0 - 15 VDC bipolar
software programmable
(16-bit DAC), max 44 mA
0 - 12 VDC bipolar
software programmable
(16-bit DAC), max 44 mA
0 - 24 V DC unipolar
2 - 30 V DC bipolar
software programmable
(16-bit DAC), max 200 mA / 2 W
TEDS
Supported, DSI® adapters
Supported, DSI® adapters
Supported, DSI® adapters
Overvoltage protection
IN+ to IN-: 50 V continuous; 200
V peak (10 msec)
IN+ to IN-: 50 V continuous; 200
V peak (10 msec)
200 V Range: 300 V;
all other ranges: 100 V (250 V
peak for 10 msec)
Counter inputs
2x Supercounter® (6 Digital In,
2 Digital out)
-
2x Supercounter® (6 Digital In,
2 Digital out)
Isolation Voltage
1000 V channel-channel,
channel-ground
Differential (ground loop
isolated)
1000 V channel-channel,
channel-ground
Input connectors
DSUB-9
CNT: 7-pin LEMO female
DSUB-9
DSUB-9
CNT: 7-pin LEMO female
Power
Power supply
9 - 36 V DC
9 - 36 V DC
9 - 36 V DC
Power consumption
Typ. 11 W (Max. 20 W)
Typ. 14 W (Max. 24 W)
Typ. 14 W (Max. 42 W,
temperature derating applies)
Environmental
Operating Temperature
-40 °C ... 60 °C
-40 °C ... 50 °C
-40 °C ... 60 °C
Storage Temperature
-40 °C ... 85 °C
-40 °C ... 85 °C
-40 °C ... 85 °C
Humidity
5 to 100 % RH
IP rating
IP65 / IP67
Shock
60 g @ 6 ms 50x, half sine, all axis
SIRIUS
®
V20-1 143 / 336

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT SIRIUS Series and is the answer not in the manual?

DEWESOFT SIRIUS Series Specifications

General IconGeneral
CategoryData Loggers
ManufacturerDEWESOFT
SeriesSIRIUS
Resolution24-bit
InterfaceUSB, EtherCAT
SoftwareDEWESoft X
ChannelsExpandable
Sampling RateUp to 200 kS/s per channel
Input TypesVoltage, Current, IEPE, Strain, Thermocouple, RTD, Resistance, Digital, Counter
Input Voltage Range±10 V, ±200 V
IsolationChannel-to-channel and channel-to-ground
ConnectivityEthernet, USB
Operating Temperature-10 °C to +50 °C
Storage Temperature-20 °C to +70 °C
Power Supply10 to 30 V DC
DimensionsVaries depending on module configuration
WeightVaries depending on module configuration
Analog InputsYes, depending on module configuration