6.1.5.3. Overall alias free bandwidth
Alias Free analog measurement bandwidth or combined filter response from Analog filter and ADC
filter response is again presented in illustration below Filtering components in SIRIUS® DualCore and
HD modules ”.
Filtering components in SIRIUS® DualCore and HD modules
Important
Analog filter (-3 dB): 
- 5 kHz up to 2 kS/s
- 100 kHz from 2 kS/s up to 204.8 kS/s
Digital filter (-3 dB): 
- 0.494 fs up to 51.2 kS/s
- 0.49 fs from 51.2 kS/s up to 102.4 kS/s
- 0.38 fs from 102.4 kS/s up to 204.8 kS/s
SIRIUS
®
V20-1 286 / 336

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT SIRIUS Series and is the answer not in the manual?

DEWESOFT SIRIUS Series Specifications

General IconGeneral
CategoryData Loggers
ManufacturerDEWESOFT
SeriesSIRIUS
Resolution24-bit
InterfaceUSB, EtherCAT
SoftwareDEWESoft X
ChannelsExpandable
Sampling RateUp to 200 kS/s per channel
Input TypesVoltage, Current, IEPE, Strain, Thermocouple, RTD, Resistance, Digital, Counter
Input Voltage Range±10 V, ±200 V
IsolationChannel-to-channel and channel-to-ground
ConnectivityEthernet, USB
Operating Temperature-10 °C to +50 °C
Storage Temperature-20 °C to +70 °C
Power Supply10 to 30 V DC
DimensionsVaries depending on module configuration
WeightVaries depending on module configuration
Analog InputsYes, depending on module configuration