EasyManuals Logo

DEWESOFT SIRIUS Series User Manual

DEWESOFT SIRIUS Series
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #278 background imageLoading...
Page #278 background image
/
SIRIUS
®
TECHNICAL REFERENCE MANUAL
Digital Filtered: the final digital FIR filter (with a sharp damping) further attenuates all signals above (f
f
),
so that we get an excellent measurement result over the whole frequency range (i.e. the effect of the
aliasing is negligible).
Important
Illustration below “Filtering components in SIRIUS® DualCore and HD modules” shows an
overview of the components in the SIRIUS® DualCore and HD modules: after the input amplifier
there is an analog filter followed by the ADC which uses oversampling and includes a Digital
ADC Filter, so that we can get excellent aliasing free measurement results.
Filtering components in SIRIUS® DualCore and HD modules
SIRIUS
®
V20-1 278 / 336

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT SIRIUS Series and is the answer not in the manual?

DEWESOFT SIRIUS Series Specifications

General IconGeneral
CategoryData Loggers
ManufacturerDEWESOFT
SeriesSIRIUS
Resolution24-bit
InterfaceUSB, EtherCAT
SoftwareDEWESoft X
ChannelsExpandable
Sampling RateUp to 200 kS/s per channel
Input TypesVoltage, Current, IEPE, Strain, Thermocouple, RTD, Resistance, Digital, Counter
Input Voltage Range±10 V, ±200 V
IsolationChannel-to-channel and channel-to-ground
ConnectivityEthernet, USB
Operating Temperature-10 °C to +50 °C
Storage Temperature-20 °C to +70 °C
Power Supply10 to 30 V DC
DimensionsVaries depending on module configuration
WeightVaries depending on module configuration
Analog InputsYes, depending on module configuration

Related product manuals