Pre-use Checkout
Instructions for Use Apollo SW 4.5n 95
Part Number: 9053586, 3rd edition
Test the O2 flush:
O
2 must be connected for the following self test.
Figure 39. Component Locations1. Occlude the Y-piece firmly onto the cone.
2. Press the >O
2+< button on the front of the
machine (1 in Figure 39).
3. Verify that the breathing bag inflates with an
audible flow.
Test the auxiliary O
2 flow meter:
1. Adjust the flow knob (2 in Figure 39) and make
sure the float moves freely over the full range of
the flow meter.
Test the function of the fresh-gas flow control
knobs:
1. Adjust the flow control knob for each available
gas (3 in Figure 39) and verify that the float
moves freely over the full range of the total flow
meter (4 in Figure 39).
Breathing bag:
1. Verify that the breathing bag is properly installed
and ready for operation (5 in Figure 39).
Verify that the vaporizers are installed and ready
for use:
Note: Before operating the vaporizer, pay special
attention to the Instructions for Use of the
vaporizer being used. Pay particular
attention to the vaporizer flow limits.
The vaporizers being used must comply with
standard ISO 8835-4 or ISO 80601-2-13. If the
internal gas measurement system fails, an
independent measurement system complying with
ISO 21647 or ISO 80601-2-55 must be used.
Note: The self test does not check for internal
vaporizer leakage; after filling or changing
vaporizers, perform the Standby leak test on
each vaporizer (see page 118).