PROline Prosonic Flow 93 FOUNDATION Fieldbus 3 Transducer Blocks (transmission blocks)
Endress+Hauser 119
Pipe Data -
Unit Pipe
Diameter
read only Displays the unit for the pipe circumference.
Note!
The unit is selected in the “System Unit - Length” parameter, see
Page 112.
Pipe Data -
Wall Thickness
AUTO - OOS Entry of the pipe wall thickness.
Measuring the wall thickness
Measurement can be carried out via:
• The local display (“Sensor” Quick Setup), see Page 19.
• The configuration and service software FieldTool.
User input:
Fixed-point number 0.1...100.0 (mm)
Factory setting:
3.2 (mm)
Note!
The unit used is displayed in the “Pipe Data - Unit Wall Thickness”
parameter.
Pipe Data -
Unit Wall
Thickness
read only Displays the unit for the thickness of the pipe wall.
Note!
The unit is selected in the “System Unit - Length” parameter, see
Page 112.
Pipe Data -
Liner Material
AUTO - OOS For selecting the material with which the pipe is lined.
Options:
LINER NONE
MORTAR
RUBBER
TAR EPOXY
OTHERS
Factory setting:
LINER NONE
Pipe Data -
Sound Velocity
Liner
AUTO - OOS
Note!
This parameter is not active unless a liner was selected in the
“Pipe Data - Liner Material” parameter.
Entry of the sound velocity of the liner.
User input:
Fixed-point number 800...6500
Factory setting:
Depends on the setting selected in the “Pipe Data - Liner Material“
parameter.
Note!
The unit used is displayed in the “Pipe Data - Unit Sound Velocity
Liner” parameter.
“TRANSDUCER_CH1” and “TRANSDUCER_CH2” (E+H parameters) /
Base index 1200 and 1300
Parameter Write access
with operating
mode
(MODE_BLK)
Description