3 Transducer Blocks (transmission blocks) PROline Prosonic Flow 93 FOUNDATION Fieldbus
126 Endress+Hauser
Sensor Param. -
Unit ARC Length
read only Displays the unit used for the display in the “Sensor Param. - ARC
Length” parameter.
Note!
The unit is selected in the “System Unit - Length” parameter, see
Page 112.
Sensor Param. -
Path Length
read only
Note!
This parameter is not active unless "INSERTION" was selected in
the “Sensor Param. - Measurement” parameter.
Displays the path length.
Display:
Max. 5-digit number
Note!
The unit used is displayed in the “Sensor Param. - Unit Path
Length” parameter.
Sensor Param. -
Unit Path Length
read only Displays the unit used for the display in the “Sensor Param. - Path
Length” parameter.
Note!
The unit is selected in the “System Unit - Length” parameter, see
Page 112.
Calibration Data -
Calibration
Factor
read only Displays the current calibration factor for the sensors.
Calibration Data -
Zero Point
AUTO - OOS For displaying or manually changing the value for the zero point
correction currently used.
User input:
5-digit floating-point number, including unit and sign
(e.g. +10.0 ns)
Calibration Data -
Correction Factor
AUTO - OOS Entry of a correction factor defined by the customer.
User input:
5-digit floating-point number
Calibration Data -
Deviation Sensor
Distance
AUTO - OOS
Note!
This parameter is not active unless "INSERTION" was selected in
the “Sensor Param. - Measurement” parameter.
Entry of the sensor distance deviation.
User input:
5-digit floating-point number
Factory setting:
0
Note!
The unit used is displayed in the “Calibration Data - Unit Deviation
Sensor Distance” parameter.
“TRANSDUCER_CH1” and “TRANSDUCER_CH2” (E+H parameters) /
Base index 1200 and 1300
Parameter Write access
with operating
mode
(MODE_BLK)
Description