PROline Prosonic Flow 93 FOUNDATION Fieldbus 3 Transducer Blocks (transmission blocks)
Endress+Hauser 127
Calibration Data -
Unit Deviation
Sensor Distance
read only Displays the unit used for the entry in the Calibration Data -
Deviation Sensor Distance parameter.
Note!
The unit is selected in the System Unit - Length parameter, see
Page 112.
Calibration Data -
Deviation ARC
Length
AUTO - OOS
Note!
This parameter is not active unless "INSERTION" was selected in
the Sensor Param. - Measurement parameter and "DUAL PATH"
was selected in the Sensor Param. - Sensor Configuration
parameter.
Entry of the arc length deviation.
User input:
5-digit floating-point number
Factory setting:
0
Note!
The unit used is displayed in the Calibration Data - Unit Deviation
ARC Length parameter.
Calibration Data -
Unit Deviation
ARC Length
read only Displays the unit used for the entry in the Calibration Data -
Deviation ARC Length parameter.
Note!
The unit is selected in the System Unit - Length parameter, see
Page 112.
Calibration Data -
Deviation Path
Length
AUTO - OOS
Note!
This parameter is not active unless "INSERTION" was selected in
the Sensor Param. - Measurement parameter.
Entry of the path length deviation.
User input:
5-digit floating-point number
Factory setting:
0
Note!
The unit used is displayed in the Calibration Data - Unit Deviation
Path Length parameter.
Calibration Data -
Unit Deviation
Path Length
read only Displays the unit used for the entry in the Calibration Data -
Deviation Path Length parameter.
Note!
The unit is selected in the System Unit - Length parameter, see
Page 112.
TRANSDUCER_CH1 and TRANSDUCER_CH2 (E+H parameters) /