Proline Prosonic Flow W 400 HART Operation
Endress+Hauser 121
Parameter Prerequisite Description User interface
Flow velocity Displays the average flow velocity that
is currently calculated.
Dependency
The unit is taken from: Velocity unit
parameter
Signed floating-point
number
Temperature Temperature is not entered as a fixed
value.
Displays the temperature that is
currently measured.
Dependency
The unit is taken from: Temperature
unit parameter
Signed floating-point
number
11.4.2 System values
The System values submenu contains all the parameters needed to display the current
measured values for every system value.
Navigation
"Diagnostics" menu → Measured values → System values
System values
Signal strength
→  121
Acceptance rate
→  121
Signal to noise ratio
→  121
Turbulence
→  121
Parameter overview with brief description
Parameter Description User interface
Signal strength Displays the current signal strength (0 to 100 dB).
Assessment of the signal strength:
< 10 dB: bad
> 90 dB: very good
Signed floating-point number
Acceptance rate Displays the ratio of the number of ultrasonic signals accepted
for flow calculation and the total number of ultrasonic signals
emitted.
0 to 100 %
Signal to noise ratio Displays the current signal to noise ratio (0 to 100 dB).
Assessment of the signal-to-noise ratio:
< 20 dB: bad
> 50 dB: very good
Signed floating-point number
Turbulence Displays the current turbulence. Signed floating-point number
11.4.3 Input values
The Input values submenu guides you systematically to the individual input values.
The submenu only appears if the device was ordered with a status input →  46..