Technical data Proline Prosonic Flow W 400 HART
168 Endress+Hauser
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
HistoROM backup T-DAT S-DAT
Available data • Event logbook, e.g. diagnostic
events
• Device firmware package
• Measured value logging ("Extended HistoROM"
order option)
• Current parameter data record (used by
firmware at run time)
• Indicator (minimum/maximum values)
• Totalizer value
• Sensor data: e.g.
• Serial number
• Device configuration (e.g. SW options,
fixed I/O or multi I/O)
Storage location Fixed on the user interface PC
board in the connection
compartment
Can be plugged into the user interface PC board in
the connection compartment
Fixed on the sensor connection board
Data backup
Automatically
• The most important device data (sensor and transmitter) are automatically saved in the
DAT modules.
• If the transmitter or measuring device is replaced: Once the T-DAT containing the
previous device data has been exchanged, the new measuring device goes into operation
immediately, without any errors.
• If the sensor is replaced: Once the S-DAT has been replaced with new device data, the
measuring device goes into operation immediately, without any errors.
Data transmission
Manual
Transfer of a device configuration to another device using the export function of the
specific operating tool, e.g. with FieldCare, DeviceCare or Web server: to duplicate the
configuration or to store in an archive (e.g. for backup purposes)
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the Extended HistoROM application package (order option) is enabled: up to 100
event messages are displayed in the events list along with a time stamp, plain text
description and remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating
tools e.g. DeviceCare, FieldCare or Web server
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Recording of 1 to 4 channels of up to 1 000 measured values (up to 250 measured values
per channel)
• User configurable recording interval
• Export the measured value log via a variety of interfaces and operating tools e.g.
FieldCare, DeviceCare or web server
16.12 Certificates and approvals
Current certificates and approvals for the product are available at www.endress.com on the
relevant product page:
1. Select the product using the filters and search field.
2. Open the product page.