Proline Prosonic Flow W 400 HART Commissioning
Endress+Hauser 111
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Security identification – Select security settings and
download these settings via
menu Data management >
Security > WLAN.
• Trusted issuer
certificate
• Device certificate
• Device private key
–
User name – Enter user name. – –
WLAN password – Enter WLAN password. – –
WLAN IP address – Enter IP address of the WLAN
interface of the device.
4 octet: 0 to 255 (in
the particular octet)
–
WLAN MAC address – Enter MAC address of the
WLAN interface of the device.
Unique 12-digit
character string
comprising letters
and numbers
Each measuring
device is given an
individual address.
WLAN passphrase The WPA2-PSK option is
selected in the Security type
parameter.
Enter the network key (8 to 32
characters).
The network key
supplied with the device
should be changed
during commissioning
for security reasons.
8 to 32-digit
character string
comprising numbers,
letters and special
characters (without
spaces)
Serial number of the
measuring device
(e.g. L100A802000)
Assign SSID name – Select which name will be used
for SSID: device tag or user-
defined name.
• Device tag
• User-defined
–
SSID name • The User-defined option is
selected in the Assign SSID
name parameter.
• The WLAN access point
option is selected in the
WLAN mode parameter.
Enter the user-defined SSID
name (max. 32 characters).
The user-defined SSID
name may only be
assigned once. If the SSID
name is assigned more
than once, the devices
can interfere with one
another.
Max. 32-digit
character string
comprising numbers,
letters and special
characters
EH_device
designation_last 7
digits of the serial
number (e.g.
EH_Prosonic_Flow_4
00_A802000)
Connection state – Displays the connection status. • Connected
• Not connected
–
Received signal strength – Shows the received signal
strength.
• Low
• Medium
• High
–
Apply changes – Use changed WLAN settings. • Cancel
• Ok
–
* Visibility depends on order options or device settings
10.5.6 Performing Heartbeat basic setup
Heartbeat setup submenu guides the user systematically through all the parameters that
can be used for the Heartbeat basic setup.
The wizard only appears if the device has the Heartbeat Verification +Monitoring
application package.