Mounting procedure Proline Prosonic Flow W 400 HART
28 Endress+Hauser
Medium pressure range
No pressure limitation For correct measurement, the static pressure of the medium must
be higher than the vapor pressure.
6.1.4 Special mounting instructions
Display guard
To ensure that the display guard can be easily opened, maintain the following
minimum head clearance: 350 mm (13.8 in)
Weather protection cover
213 (8.4)
39 (1.5)
243 (9.6)
203 (8.0)
A0029552
 15 Weather protection cover; engineering unit mm (in)
6.2 Mounting the measuring device
6.2.1 Required tools
For transmitter
Torque wrench
For wall mounting:
Open-ended wrench for hexagonal screw max. M5
For pipe mounting:
Open-ended wrench AF 8
Phillips head screwdriver PH 2
For sensor
For mounting on the measuring tube: Use a suitable mounting tool.
6.2.2 Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove stick-on label on the electronics compartment cover.