Mounting procedure Proline Prosonic Flow W 400 HART
28 Endress+Hauser
Medium pressure range
No pressure limitation For correct measurement, the static pressure of the medium must
be higher than the vapor pressure.
6.1.4 Special mounting instructions
Display guard
‣
To ensure that the display guard can be easily opened, maintain the following
minimum head clearance: 350 mm (13.8 in)
Weather protection cover
213 (8.4)
39 (1.5)
243 (9.6)
203 (8.0)
A0029552
15 Weather protection cover; engineering unit mm (in)
6.2 Mounting the measuring device
6.2.1 Required tools
For transmitter
• Torque wrench
• For wall mounting:
Open-ended wrench for hexagonal screw max. M5
• For pipe mounting:
• Open-ended wrench AF 8
• Phillips head screwdriver PH 2
For sensor
For mounting on the measuring tube: Use a suitable mounting tool.
6.2.2 Preparing the measuring device
1. Remove all remaining transport packaging.
2. Remove stick-on label on the electronics compartment cover.