Commissioning Proline Prosonic Flow W 400 HART
100 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign flow direction check The Switch option is
selected in the Operating
mode parameter.
The Flow direction check
option is selected in the
Switch output function
parameter.
Select process variable for flow
direction monitoring.
Assign status The Switch option is
selected in Operating mode
parameter.
The Status option is
selected in Switch output
function parameter.
Select the device function for
which to report the status. If
the function is triggered, the
output is closed and conductive
(standard configuration).
Off
Low flow cut off
Switch-on value The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Enter limit value for switch-on
point (process variable >
switch-on value = closed,
conductive).
Signed floating-point
number
Depends on country
Switch-off value The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Enter limit value for switch-off
point (process variable <
switch-off value = open,
nonconductive).
Signed floating-point
number
Depends on country
Switch-on delay The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Enter a delay before the output
is switched on.
0.0 to 100.0 s
Switch-off delay The Switch option is
selected in the Operating
mode parameter.
The Limit option is selected
in the Switch output
function parameter.
Enter a delay before the output
is switched off.
0.0 to 100.0 s
Failure mode Select output behavior in the
event of a device alarm.
Actual status
Open
Closed
Invert output signal Invert the output signal. No
Yes
* Visibility depends on order options or device settings
10.4.8 Configuring the local display
The Display wizard guides you systematically through all the parameters that can
configured for configuring the local display.
Navigation
"Setup" menu → Display
Display
Format display
→  101