Proline Prosonic Flow W 400 HART Technical data
Endress+Hauser 155
HART protocol
The measured values are written from the automation system to the measuring device via
the HART protocol. The temperature and density measuring device must support the
following protocol-specific functions:
HART protocol
Burst mode
Status input
Maximum input values DC 30 V
6 mA
Response time Configurable: 5 to 200 ms
Input signal level Low signal (low): DC –3 to +5 V
High signal (high): DC 12 to 30 V
Assignable functions Off
Reset totalizers 1-3 separately
Reset all totalizers
Flow override
16.4 Output
Output signal Current output
Current output Can be set as:
4 to 20 mA NAMUR
4 to 20 mA US
4 to 20 mA HART
0 to 20 mA
Maximum output values DC 24 V (when idle)
22.5 mA
Load 250 to 700 Ω
Resolution 0.38 µA
Damping Configurable: 0 to 999.9 s
Assignable measured
variables
Volume flow
Mass flow
Sound velocity
Flow velocity
Electronics temperature
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Function With the order code for "Output; Input", option H: output 2 can be set as a pulse
or frequency output
With the order code for "Output; Input", option I: output 2 and 3 can be set as a
pulse, frequency or switch output
Version Passive, open collector
Maximum input values DC 30 V
250 mA
Voltage drop At 25 mA: ≤ DC 2 V
Pulse output
Pulse width Configurable: 0.05 to 2 000 ms