EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow W 400 - Environment; Mounting

Endress+Hauser Proline Prosonic Flow W 400
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data Proline Prosonic Flow W 400 HART
162 Endress+Hauser
16.7 Mounting
Mounting requirements →  19
16.8 Environment
Ambient temperature
range
→  27
Storage temperature The storage temperature for all components (except display modules and order code for
"Sensor version", options AG, AH) corresponds to the ambient temperature range→  27.
Display modules
–40 to +60 °C (–40 to +140 °F)
Relative humidity The device is suitable for use outdoors and indoors with a relative humidity of 5 to 95 %.
Operating height According to EN 61010-1
≤ 2 000 m (6 562 ft)
> 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW
Series)
Degree of protection Transmitter
IP66/67, type 4X enclosure, suitable for pollution degree 4
When the housing is open: IP20, type 1 enclosure, suitable for pollution degree 2
Display module: IP20, type 1 enclosure, suitable for pollution degree 2
Sensor
Standard: IP66/67, type 4X enclosure, suitable for pollution degree 4
Optionally available: IP68, type 6P enclosure, suitable for pollution degree 4
External WLAN antenna
IP67
Shock and vibration
resistance
Vibration sinusoidal, in accordance with IEC 60068-2-6
2 to 8.4 Hz, 7.5 mm peak
8.4 to 2 000 Hz, 2 g peak for transmitter, 1 g peak for sensor
Vibration broad-band random, according to IEC 60068-2-64
10 to 200 Hz, 0.01 g
2
/Hz
200 to 2 000 Hz, 0.003 g
2
/Hz
Total: 2.70 g rms
Shock half-sine, according to IEC 60068-2-27
6 ms 50 g
Rough handling shocks according to IEC 60068-2-31

Table of Contents