EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow W 400 - Page 161

Endress+Hauser Proline Prosonic Flow W 400
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow W 400 HART Technical data
Endress+Hauser 161
Example of max. measurement error (volume flow)
0
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
[%]
2 4 6 8 10 12 14
2
1
[m/s]
[ft/s]
0 2010 30
40
50
A0041973
 56 Example of max. measurement error (volume flow) in % o.r.
1 Pipe diameter < DN 100 (4")
2 Pipe diameter ≥ DN 100 (4")
Accuracy of outputs
The outputs have the following base accuracy specifications.
Current output
Accuracy Max. ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range)
Repeatability o.r. = of reading
±0.3% for flow velocities >0.3 m/s (1 ft/s)
Influence of ambient
temperature
Current output
o.r. = of reading
Temperature coefficient Max. ±0.005 % o.r./°C
Pulse/frequency output
Temperature coefficient No additional effect. Included in accuracy.

Table of Contents