Commissioning Proline Prosonic Flow W 400 HART
86 Endress+Hauser
Parameter overview with brief description
Parameter Description Selection Factory setting
Volume flow unit Select volume flow unit.
Result
The selected unit applies to:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
m³/h
ft³/min
Volume unit Select volume unit. Unit choose list Country-specific:
ft³
Mass flow unit Select mass flow unit.
Result
The selected unit applies to:
Output
Low flow cut off
Simulation process variable
Unit choose list Country-specific:
kg/h
lb/min
Mass unit Select mass unit. Unit choose list Country-specific:
kg
lb
Velocity unit Select velocity unit.
Effect
The selected unit applies to:
Flow velocity
Sound velocity
Unit choose list Depends on country:
m/s
ft/s
Temperature unit Select temperature unit.
Result
The selected unit applies to:
Temperature
Electronic temperature parameter
(6053)
External temperature parameter (6080)
Reference temperature parameter
(1816)
Unit choose list Country-specific:
°C
°F
Density unit Select density unit.
Result
The selected unit applies to:
Output
Simulation process variable
Unit choose list Country-specific:
kg/dm³
lb/ft³
Length unit Select the unit of length. Unit choose list Country-specific:
mm
in
10.4.3 Configuring the measuring point
The "Measuring point " wizard guides you systematically through all the parameters that
must be set for the configuration of the measuring point.
Navigation
"Setup" menu → Measuring point
Measuring point
Measuring point configuration
→  88