Technical data Proline Prosonic Flow W 400 HART
158 Endress+Hauser
Galvanic isolation The following connections are galvanically isolated from each other:
Inputs
Outputs
Power supply
DN 50 to 4000 (2 to 160") and non-hazardous area: The clamp-on sensors can also be
mounted on cathodically protected pipes. Solution available on request.
Protocol-specific data HART
For information on the device description files
For information on the dynamic variables and measured variables (HART device
variables) →  79
16.5 Power supply
Terminal assignment →  46
Supply voltage Transmitter
Order code for "Power supply" terminal voltage Frequency range
Option L
DC 24 V ±25%
AC 24 V ±25% 50/60 Hz, ±4 Hz
AC 100 to 240 V –15 to +10% 50/60 Hz, ±4 Hz
Power consumption
Order code for "Output" Maximum power consumption
Option H: 4-20mA HART, pulse/frequency output,
switch output
30 VA/8 W
Option I: 4-20mA HART, 2 x pulse/frequency/switch
output, status input
30 VA/8 W
Current consumption Transmitter
Order code for "Power supply" Maximum
Current consumption
Maximum
switch-on current
Option L: AC 100 to 240 V 145 mA 25 A (< 5 ms)
Option L: AC/DC 24 V 350 mA 27 A (< 5 ms)
Device fuse Fine-wire fuse (slow-blow):
DC 24 V: T1A
AC 100 to 240 V: T1A
Power supply failure Totalizers stop at the last value measured.
Depending on the device version, the configuration is retained in the device memory or
in the pluggable data memory (HistoROM DAT).
Error messages (incl. total operated hours) are stored.