Proline Prosonic Flow W 400 HART Diagnosis and troubleshooting
Endress+Hauser 141
Diagnostic
number
Short text Remedy instructions Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of process
803 Loop current 1 faulty 1. Check wiring
2. Change I/O module
F Alarm
832 Electronics temperature
too high
Reduce ambient temperature S Warning
1)
833 Electronics temperature
too low
Increase ambient temperature S Warning
1)
841 Flow velocity too high Reduce flow rate S Warning
1)
842 Process value below limit Low flow cut off active!
Check low flow cut off configuration
S Warning
1)
870 Measuring inaccuracy
increased
1. Check process
2. Increase flow volume
F Alarm
1)
881 Signal to noise ratio too
low
1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
F Alarm
882 Input signal faulty 1. Check input signal
parameterization
2. Check external device
3. Check process conditions
F Alarm
930 Sound velocity too high 1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
S Warning
1)
931 Sound velocity too low 1. Check process conditions
2. Clean/repl. transd. (inline) / check
sensor pos. and coupling (clamp on)
3. Replace sensor electronic module
(ISEM)
S Warning
1)
953 Asymmetry noise signal
too high path 1 to n
1. Check process conditions
2. Clean or replace transducers
3. Replace sensor electronic module
(ISEM)
M Alarm
1) Diagnostic behavior can be changed.
12.8 Pending diagnostic events
The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
Via local display →  134
Via web browser →  135
Via "FieldCare" operating tool →  137
Via "DeviceCare" operating tool →  137
Other pending diagnostic events can be displayed in the Diagnostic list submenu
→  142.