Proline Prosonic Flow W 400 HART Mounting procedure
Endress+Hauser 29
6.2.3 Mounting the sensor
L
WARNING
Risk of injury when mounting sensors and strapping bands!
‣
Suitable gloves and goggles must be worn due to the increased risk of cuts.
Sensor configuration and settings
DN 15 to 65 (½ to 2½") DN 50 to 4000 (2 to 160")
Strapping band Strapping band Welded bolt
2 traverses
[mm (in)]
1 traverse
[mm (in)]
2 traverses
[mm (in)]
1 traverse
[mm (in)]
2 traverses
[mm (in)]
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
Sensor distance
1)
– Wire length → 38 Measuring rail
1) 2)
Wire length Measuring rail
1) 2)
1) Depends on the conditions at the measuring point (e.g. measuring pipe, medium). The dimension can be
determined via FieldCare or Applicator. See also Result sensor distance / measuring aid parameter in
Measuring point submenu
2) Up to DN 600 (24")
Determining the mounting positions of the sensor
Sensor holder with U-shaped screws )
Can be used for
• Measuring devices with measuring range DN 15 to 65 (½ to 2½")
• Mounting on pipes DN 15 to 32 (½ to 1¼")
Procedure:
1. Disconnect the sensor from the sensor holder.
2. Position the sensor holder on the measuring pipe.
3. Insert the U-shaped screws through the sensor holder and lightly grease the threads.
4. Screw the nuts onto the U-shaped screws.
5. Position the sensor holder exactly and tighten the nuts evenly.
A0043369
16 Holder with U-shaped screws
1 Sensor holder