EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow W 400 - Page 38

Endress+Hauser Proline Prosonic Flow W 400
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mounting procedure Proline Prosonic Flow W 400 HART
38 Endress+Hauser
Procedure with a tape measure:
1. Use a tape measure to determine the pipe diameter d.
2. Mount the opposite mounting bolt at d/2 from the front mounting bolt. The distance
must be d/2 = d'/2 on both sides.
3. Check distance B.
d/2 = d‘/2
B
d/2
d‘/2
A0052445
 28 Positioning the strapping bands and mounting bolts with a tape measure (steps 2 to 4)
Fastening the sensors:
1. Fit the sensor holders over the individual mounting bolts and tighten securely with
the locking nut.
13 mm
A0043381
 29 Mounting the sensor holders
2. Stick the coupling pad under the sensor →  172. Alternatively, coat the contact
surfaces of the sensor evenly with coupling gel (approx. 1 mm (0.04 in)). In doing so,
start from the groove through the center to the opposite edge.
A0043382
 30 Coating the contact surfaces of the sensor with coupling gel (if there is no coupling pad)
3. Insert the sensor into the sensor holder.

Table of Contents

Related product manuals