Technical data Proline Prosonic Flow W 400 HART
156 Endress+Hauser
Maximum pulse rate 10 000 Impulse/s
Pulse value Configurable
Assignable measured
variables
Volume flow
Mass flow
Frequency output
Output frequency Configurable: 0 to 12 500 Hz
Damping Configurable: 0 to 999 s
Pulse/pause ratio 1:1
Assignable measured
variables
Volume flow
Mass flow
Sound velocity
Flow velocity
Electronics temperature
Switch output
Switching behavior Binary, conductive or non-conductive
Switching delay Configurable: 0 to 100 s
Number of switching
cycles
Unlimited
Assignable functions Off
On
Diagnostic behavior
Limit value
Volume flow
Mass flow
Sound velocity
Flow velocity
Totalizer 1-3
Electronics temperature
Flow direction monitoring
Status
Low flow cut off
Signal on alarm Depending on the interface, failure information is displayed as follows:
Current output 4 to 20 mA
4 to 20 mA
Failure mode Choose from:
4 to 20 mA in accordance with NAMUR recommendation NE 43
4 to 20 mA in accordance with US
Min. value: 3.59 mA
Max. value: 22.5 mA
Definable value between: 3.59 to 22.5 mA
Actual value
Last valid value
0 to 20 mA
Failure mode Choose from:
Max. alarm: 22 mA
Definable value between: 0 to 22.5 mA
HART current output
Device diagnostics Device condition can be read out via HART Command 48