Proline Prosonic Flow W 400 HART Technical data
Endress+Hauser 167
Supported operating
tools
Operating unit Interface Additional information
Field Xpert SMT70/77/50 All Fieldbus protocols
WLAN interface
Bluetooth
CDI-RJ45 service
interface
Operating Instructions
BA01202S
Device description files:
Use update function of
handheld terminal
SmartBlue app Smart phone or tablet
with iOs or Android
WLAN →  153
Other operating tools based on FDT technology with a device driver such as DTM/
iDTM or DD/EDD can be used for device operation. These operating tools are available
from the individual manufacturers. Integration into the following operating tools,
among others, is supported:
FactoryTalk AssetCentre (FTAC) from Rockwell Automation →
www.rockwellautomation.com
Process Device Manager (PDM) from Siemens → www.siemens.com
Asset Management Solutions (AMS) from Emerson → www.emersonprocess.com
FieldCommunicator 375/475 from Emerson → www.emersonprocess.com
Field Device Manager (FDM) from Honeywell → www.process.honeywell.com
FieldMate from Yokogawa → www.yokogawa.com
PACTWare → www.pactware.com
The related device description files are available: www.endress.com → Download Area
Web server
With the integrated web server, the device can be operated and configured via a web
browser service interface (CDI-RJ45) or WLAN interface. The structure of the operating
menu is the same as for the local display. In addition to the measured values, status
information on the device is displayed and can be used to monitor device health.
Furthermore the device data can be managed and the network parameters can be
configured.
A device that has a WLAN interface (can be ordered as an option) is required for the
WLAN connection: order code for "Display", option G "4-line illuminated; touch control
+WLAN". The device acts as an access point and enables communication by computer or a
mobile handheld terminal.
Supported functions
Data exchange between the operating unit (such as a notebook, for example,) and
measuring device:
Upload the configuration from the measuring device (XML format, configuration
backup)
Save the configuration to the measuring device (XML format, restore configuration)
Export event list (.csv file)
Export parameter settings (.csv file or PDF file, document the measuring point
configuration)
Export the Heartbeat verification report (PDF file, only available with the Heartbeat
Verification →  170 application package)
Flash firmware version for device firmware upgrade, for example
Download driver for system integration
Visualize up to 1000 saved measured values (only available with the Extended
HistoROM application package→  170)
HistoROM
data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.