Proline Prowirl R 200 HART Technical data
Endress+Hauser 189
1000 10000 100000
0
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
[%]
V / D
i
³
A0034414
31
Repeatability = 0.1 % o.r. at a measured volume [m
3
] of V = 1000 ⋅ D
i
3
The repeatability can be improved if the measured volume is increased. Repeatability is not
a device characteristic but a statistical variable that is dependent on the boundary
conditions indicated.
Response time If all the configurable functions for filter times (flow damping, display damping, current
output time constant, frequency output time constant, status output time constant) are set
to 0, in the event of vortex frequencies of 10 Hz and higher a response time of max(T
ν
,
100 ms) can be expected.
In the event of measuring frequencies < 10 Hz, the response time is > 100 ms and can be
up to 10 s. T
ν
is the average vortex period duration of the flowing fluid.
Influence of ambient
temperature
Current output
o.r. = of reading
Additional error, in relation to the span of 16 mA:
Temperature coefficient at
zero point (4 mA)
0.02 %/10 K
Temperature coefficient
with span (20 mA)
0.05 %/10 K
Pulse/frequency output
o.r. = of reading
Temperature coefficient Max. ±100 ppm o.r.
16.7 Installation
"Mounting requirements" → 22
16.8 Environment
Ambient temperature
range