Loading...
Commissioning Proline Prowirl R 200 HART
120 Endress+Hauser
Options Description
Duplicate The transmitter configuration from another device is duplicated to the device using
the display module.
Clear backup data The backup copy of the device configuration is deleted from the display module of
the device.
HistoROM backup
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.5.7 Using parameters for device administration
The Administration submenu systematically guides the user through all the parameters
that can be used for device administration purposes.
Navigation
"Setup" menu → Advanced setup → Administration
Administration
Define access code
Define access code
→  120
Confirm access code
→  120
Device reset
→  120
Parameter overview with brief description
Parameter Description User entry / Selection Factory setting
Define access code Restrict write-access to parameters to
protect the configuration of the device
against unintentional changes via the local
display.
0 to 9 999 0
Confirm access code Confirm the entered access code. 0 to 9 999 0
Device reset Reset the device configuration - either
entirely or in part - to a defined state.
Cancel
To factory defaults
To delivery settings
Restart device
Cancel
10.6 Configuration management
After commissioning, you can save the current device configuration, copy it to another
measuring point or restore the previous device configuration.
You can do so using the Configuration management parameter and the related options
found in the Configuration backup display submenu.