Commissioning Proline Prowirl R 200 HART
124 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry
Factory setting
Pulse output simulation In the Operating mode
parameter, the Pulse option is
selected.
Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
(→ 86) defines the
pulse width of the pulses
output.
• Off
• Fixed value
• Down-counting
value
Off
Pulse value In the Pulse output
simulation parameter
(→ 124), the Down-
counting value option is
selected.
Enter the number of pulses for
simulation.
0 to 65 535 0
Switch output simulation In the Operating mode
parameter, the Switch option
is selected.
Switch the simulation of the
switch output on and off.
• Off
• On
Off
Switch status In the Switch output
simulation parameter
(→ 124) Switch output
simulation 1 to n parameter
Switch output simulation
1 to n parameter, the On
option is selected.
Select the status of the status
output for the simulation.
• Open
• Closed
Open
Device alarm simulation – Switch the device alarm on and
off.
• Off
• On
Off
Diagnostic event category – Select a diagnostic event
category.
• Sensor
• Electronics
• Configuration
• Process
Process
Diagnostic event simulation – Select a diagnostic event to
simulate this event.
• Off
• Diagnostic event
picklist (depends
on the category
selected)
Off
* Visibility depends on order options or device settings
10.8 Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
• Write protection via access code
• Write protection via write protection switch
• Write protection via keypad lock
10.8.1 Write protection via access code
The effects of the user-specific access code are as follows:
• Via local operation, the parameters for the measuring device configuration are write-
protected and their values can no longer be changed.
• Device access is protected via the Web browser, as are the parameters for the measuring
device configuration.
Defining the access code via local display
1. Navigate to the Enter access code parameter.
2. Define a max. 16-digit character string comprising numbers, letters and special
characters as the access code.