Commissioning Proline Prowirl R 200 HART
84 Endress+Hauser
10.4.5 Configuring the current output
The Current output wizard guides you systematically through all the parameters that have
to be set for configuring the current output.
Navigation
"Setup" menu → Current output 1 to n
‣
Current output 1 to n
Assign current output 1 to n
→ 84
Current span
→ 84
4 mA value
→ 84
20 mA value
→ 85
Fixed current
→ 85
Damping output 1 to n
Failure mode
→ 85
Failure current
→ 85
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Assign current output – Select process variable for
current output.
• Off
• Volume flow
• Corrected volume
flow
• Mass flow
• Flow velocity
• Temperature
• Pressure
• Calculated
saturated steam
pressure
*
• Total mass flow
*
• Energy flow
*
• Heat flow
difference
*
Volume flow
Current span – Select current range for
process value output and
upper/lower level for alarm
signal.
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• Fixed current
Country-specific:
• 4...20 mA NAMUR
• 4...20 mA US
4 mA value One of the following options is
selected in the Current span
parameter (→ 84):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
Enter 4 mA value. Signed floating-point
number
Country-specific:
• 0 m³/h
• 0 ft³/min