Technical data Proline Prowirl R 200 HART
202 Endress+Hauser
Via remote display FHX50
• The remote display FHX50 can be ordered as an optional extra→ 169.
• The remote display FHX50 cannot be combined with the order code for "Sensor
version; DSC sensor; measuring tube", option DA "mass steam" or option DB "mass
gas/liquid".
A0032215
33 FHX50 operating options
1 SD02 display and operating module, push buttons: cover must be opened for operation
2 SD03 display and operating module, optical buttons: operation possible through cover glass
Display and operating elements
The display and operating elements correspond to those of the display module .
Remote operation → 64
Service interface → 65
16.12 Certificates and approvals
CE mark The measuring system is in conformity with the statutory requirements of the applicable
EU Directives. These are listed in the corresponding EU Declaration of Conformity along
with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications
and Media Authority (ACMA)".
Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this
document on the nameplate.
Functional safety The measuring device can be used for flow monitoring systems (min., max., range) up to
SIL 2 (single-channel architecture; order code for "Additional approval", option LA) and SIL
3 (multichannel architecture with homogeneous redundancy) and is independently
evaluated and certified by the TÜV in accordance with IEC 61508.