Proline Prowirl R 200 HART Technical data
Endress+Hauser 183
Protocol-specific data
Manufacturer ID 0x11
Device type ID 0x0038
HART protocol revision 7
Device description files
(DTM, DD)
Information and files under:
www.endress.com
HART load Min. 250 Ω
Max. 500 Ω
System integration For information on system integration, see .→  69
Measured variables via HART protocol
Burst Mode functionality
16.5 Power supply
Terminal assignment →  37
Supply voltage Transmitter
An external power supply is required for each output.
The following supply voltage values apply for the outputs available:
Supply voltage for a compact version without a local display
1)
Order code for "Output; input" Minimum
terminal voltage
2)
Maximum
terminal voltage
Option A: 4-20 mA HART ≥ DC 12 V DC 35 V
Option B: 4-20 mA HART, pulse/
frequency/switch output
≥ DC 12 V DC 35 V
Option C: 4-20 mA HART + 4-20 mA
analog
≥ DC 12 V DC 30 V
Option D: 4-20 mA HART, pulse/
frequency/switch output, 4-20 mA current
input
3)
≥ DC 12 V DC 35 V
1) In event of external supply voltage of the power supply unit with load
2) The minimum terminal voltage increases if local operation is used: see the following table
3) Voltage drop 2.2 to 3 V for 3.59 to 22 mA
Increase in minimum terminal voltage
Order code for "Display; operation"
Increase in minimum
terminal voltage
Option C:
Local operation SD02
+ DC 1 V
Option E:
Local operation SD03 with lighting
(backlighting not used)
+ DC 1 V
Option E:
Local operation SD03 with lighting
(backlighting used)
+ DC 3 V