Proline Prowirl R 200 HART Commissioning
Endress+Hauser 119
You can do so using the Configuration management parameter and the related options
found in the Configuration backup display submenu.
Navigation
"Setup" menu → Advanced setup → Configuration backup display
‣
Configuration backup display
Operating time
→ 119
Last backup
→ 119
Configuration management
→ 119
Comparison result
→ 119
Parameter overview with brief description
Parameter Prerequisite Description User interface /
Selection
Factory setting
Operating time – Indicates how long the device
has been in operation.
Days (d), hours (h),
minutes (m) and
seconds (s)
–
Last backup A local display is provided. Indicates when the last data
backup was saved to the
display module.
Days (d), hours (h),
minutes (m) and
seconds (s)
–
Configuration management A local display is provided. Select action for managing the
device data in the display
module.
• Cancel
• Execute backup
• Restore
• Duplicate
• Compare
• Clear backup data
Cancel
Comparison result A local display is provided. Comparison between present
device data and display
backup.
• Settings identical
• Settings not
identical
• No backup
available
• Backup settings
corrupt
• Check not done
• Dataset
incompatible
Check not done
Function scope of the "Configuration management" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
Execute backup A backup copy of the current device configuration is saved from the HistoROM
backup to the display module of the device. The backup copy includes the
transmitter data of the device.
Restore The last backup copy of the device configuration is restored from the display
module to the device's HistoROM backup. The backup copy includes the transmitter
data of the device.
Compare The device configuration saved in the display module is compared with the current
device configuration of the HistoROM backup.