Proline Prowirl R 200 HART Commissioning
Endress+Hauser 81
Parameter Prerequisite Description Selection / User
entry
Factory setting
Gas type The following conditions are
met:
In the Select medium
parameter, the Gas option is
selected.
In the Select gas type
parameter, the Single gas
option is selected.
Select measured gas type. Hydrogen H2
Helium He
Neon Ne
Argon Ar
Krypton Kr
Xenon Xe
Nitrogen N2
Oxygen O2
Chlorine Cl2
Ammonia NH3
Carbon monoxide
CO
Carbon dioxide
CO2
Sulfur dioxide SO2
Hydrogen sulfide
H2S
Hydrogen chloride
HCl
Methane CH4
Ethane C2H6
Propane C3H8
Butane C4H10
Ethylene C2H4
Vinyl Chloride
C2H3Cl
Methane CH4
Relative humidity The following conditions are
met:
In the Select medium
parameter, the Gas option is
selected.
In the Select gas type
parameter, the Air option is
selected.
Enter humidity content of air
in %.
0 to 100 % 0 %
Steam calculation mode The Steam option is selected
in the Select medium
parameter parameter.
Select calculation mode of
steam: based on saturated
steam (T-compensated) or
automatic detection (p-/T-
compensated).
Saturated steam
(T-compensated)
Automatic (p-/T-
compensated)
Saturated steam (T-
compensated)
Select liquid type The following conditions are
met:
Order code for "Sensor
version",
Option "Mass (integrated
temperature
measurement)"
or
Option "Mass (integrated
pressure/temperature
measurement)"
The Liquid option is
selected in the Select
medium parameter
parameter.
Select measured liquid type. Water
LPG (Liquefied
Petroleum Gas)
User-specific liquid
Water