9 Alert Information and FAQ
109
FOCUSED PHOTONICS INC
9 Alert Information and FAQ
9.1 Alert Information
Alert Code Content Level Measures (Mainly for Protection)
01 Mismatch of sensor type and
pixel
Failure The spectrometer already runs at
default configuration. You need to
re-configure the light source byte
02 Over limit of optical channel
1
Failure The spectrometer already runs at
default configuration. You need to
re-configure the optical channel 1
03 Over limit of optical channel
2
Failure The spectrometer already runs at
default configuration. You need to
re-configure the optical channel 2
04 EEPROM error Failure Restart the system. If the problem
still cannot be solved, you should
change the EEPROM
05 Step motor error Failure Restart the instrument. If the
problem still cannot be solved,
please check the relevant circuit of
the motor, the motor and the optical
coupling.
06 Error in EEPROM’s vector
index
Failure The vector table should be
configured
11 Error in display board
E2PROM
Failure
12 Error in display board RTC Failure
13 Gas room temperature over
limit
Failure
14 Gas pipe jammed Failure
15 Temperature control error Failure
16 Pressure error Failure
17 SO2 density alert Warning
18 NO density alert Warning
19 O2 density alert Warning