EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 4-19 Spectrometer Parameter Setup; Fig 4-20 Checking Spectrum Strength

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operations of OMA-2000 Spectroscopic Gas Analyzer
42
FOCUSED PHOTONICS INC
¾ Xenon light pulse times: Set the pulse times for the xenon light within the range of
1~10.
Fig 6-20 Spectrometer Parameter Setup
z Spectrum strength: Check the spectrum strength of a pixel, as shown in Fig 6-21.
Fig 6-21 Checking Spectrum Strength
Please follow the steps below to check the spectrum strength:
Step 12 Press the “F2” button and set the pixel using the number buttons within the range
of (1~511);
Step 13 Press the “Enter” button to confirm the setup and press the “F3” button to read the
spectrum strength of the pixel.
z Pressure: Display the current pressure and pressure input method in the pressure room,
as shown in Fig 6-22.

Table of Contents