EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 4-13 Baud Rate Setup

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operations of OMA-2000 Spectroscopic Gas Analyzer
38
FOCUSED PHOTONICS INC
z Instrument address: The communication address of the instrument for communication
with the PC.
Please follow the steps below to set the baud rate:
Step 5 Press the “F1” button on the communication setup page in Fig 6-13 to enter the
baud rate setup, as shown in Fig 6-14.
Fig 6-14 Baud Rate Setup
Step 6 In the baud rate setup page, you can press the “F1~F3” buttons to select the baud
rate (9600, 19200 or 57600) for communication between the instrument and the
PC. It is recommended to select 57600.
Step 7 After the baud rate is selected, please return to the communication setup page in
Fig 6-13. Press “F3” button to confirm the setup.
Please follow the steps below to setup the instrument address:
Step 8 On the communication setup page as shown in Fig 6-13, press the “F2” button to
enter the instrument address setup, as shown in Fig 6-15;
Step 9 On the communication setup page, press the number buttons to input the address
within the range of 1~247 for communication between the instrument and the PC;
Step 10 Press the “Enter” button to confirm the setup;
Step 11 After the instrument address has been configured, please return to the
communication setup page and press the “F3” button to activate the settings.

Table of Contents