6 Operations of OMA-2000 Spectroscopic Gas Analyzer
41
FOCUSED PHOTONICS INC
6.5.5 Parameter Setup
The parameter setup is the 4
th
item of the instrument setup menu. Press the “F4” button to
enter the parameter setup page as shown in Fig 6-18 and Fig 6-19.
Fig 6-18 Parameter Setup 1
Fig 6-19 Parameter Setup 2
The meanings of items on the parameter setup page are:
z Spectrometer: The parameter setup related to the spectrometer measurement:
¾ Xenon voltage supply: Set the voltage for xenon light within the range of (2.5~5)v;
¾ Spectrum average times: Set the average times of the spectrum within the range of
1~100;