EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 5-23 System Constant Configuration

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Operations of CEMS2000 Monitoring Software
72
FOCUSED PHOTONICS INC
5.2.2.1 System Constant Configuration
This function is used for configuring the system constants, as shown in Fig 7-22.
Fig 7-22 System Constant Configuration
In the constant configuration window, you can input the corresponding value of the constant
in the “Constant Value” box. Click the “Save” button to save the constant and click the “Close”
button to return to the main window.
Note:
You should configure the measurement constant for each measuring point. Normally you can
use the default settings.
7.2.2.2 Configuration of Default Measurement Range
This function is used for configuring the measurement range of the measuring variables
displayed on the PC.

Table of Contents