4 System Composition
10
FOCUSED PHOTONICS INC
z Material: WZP PT resistance
z Range: (0300) (optional according to actual conditions)
z Accuracy: ±0.5%
z Input power: 24V DC
z Output current: Two wire (4-20)mA
z Inserted length: (200-2000)mm or customized
z Others: The bushing of temperature transmitter is covered with anti-corrosion material
(PTFE) and completed with installation base.
Fig 4-6 Exterior of Temperature Transmitter
4.3.2 Pressure Transmitter
z Instrument: Pressure transmitter
z Model: STP
z Principle: Pressure sensor
z Range: (-10~10)Kpa, (optional according to actual conditions)
z Accuracy: ±0.5%
z Input power: 24V DC
z Output current: Two wire (4-20)mA
4.3.3 Flow Rate Transmitter
z Instrument: Flow rate meter
z Model: PT-1
z Principle: Pitot tube and pressure difference transmitter 3051CD1
z Range: (0~10)m/s to (0~80)m/s, optional according to actual conditions
z Accuracy: ±5%F.S.
z Output power: 24V DC
z Output current: Two wire (4-20)mA
z Inserted length: (200-2000)mm or customized
z Others: The surface is covered with anti-corrosion material (PTFE) and completed with
installation flange, bolts and washers.
z Purging system: Includes SMC electro-magnetic valve, air tap and three-way valve.
z Input voltage: 220V AC