EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 4-34 Zero Adjustment; Calibration Setup; Fig 4-35 Zero Adjustment

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operations of OMA-2000 Spectroscopic Gas Analyzer
52
FOCUSED PHOTONICS INC
Fig 6-35 Zero Adjustment / Calibration Setup
The meanings of items on this page are:
z Calibration density: Set the density of the standard gas within the range of (0~50000)
ppm
z Manual zero adjustment: Perform manual zero adjustment operations.
Please follow the steps below for manual zero adjustment:
Step 16 Press the “F2” button to enter the manual zero adjustment page, as shown in Fig
6-36.
Fig 6-36 Zero Adjustment
Step 17 Press the “Enter” button on the page of zero adjustment to start adjustment, as
shown in Fig 6-37.

Table of Contents