EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 4-2 Main Page (Input Password)

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operations of OMA-2000 Spectroscopic Gas Analyzer
29
FOCUSED PHOTONICS INC
Fig 6-2 Main Page
Here we take the “Change Password” for example to introduce the usage of above buttons:
Step 1 Press the “Set” button on the main page. The system will automatically display the
password input dialog box, as shown in Fig 6-3.
Fig 6-3 Main Page (Input Password)
Step 2 Input the password using number buttons in the dialog box shown in Fig 6-3. The
default password is 111111. After the password is input correctly, press the “Enter”
button to enter instrument setup;
Step 3 On the instrument setup page, press the “F5” button to enter the next page of setup
and press the “F2” button to enter the changing password page, as shown in Fig
6-4.

Table of Contents