EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 4-36 Zero Adjustment in Progress; Fig 4-37 Calibration

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Operations of OMA-2000 Spectroscopic Gas Analyzer
53
FOCUSED PHOTONICS INC
Fig 6-37 Zero Adjustment in Progress
Step 18 During the adjustment, you can press the “ESC” button to cancel the progress.
After the adjustment is finished, the system will return to the page as shown in Fig
6-36 but the prompt of “Start adjustment?” is be available.
z Manual calibration: Perform the manual calibration operations.
Please follow the steps below to perform manual calibration:
Step 19 Press the “F3” button on the zero adjustment / calibration page as shown in Fig
6-35 to enter the manual calibration, as shown in Fig 6-38.
Fig 6-38 Calibration
Step 20 Press the “Enter” button on the calibration page to start the calibration, as shown in
Fig 6-39.

Table of Contents