EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 3-13 Zero Adjustment Page on OMA-2000 Analyzer; Fig 3-14 Manual Calibration

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 System Operations
22
Counter-clockwise rotate the knob to open the SO
2
gas bottle and slowly adjust the
“two-level pressure adjuster” to a proper pressure value. Normally the “low
pressure meter” should be maintained around “0.2” MPa;
Step 2 Click the SO
2
composition management page to enter the calibration page as
shown in Fig 5-13.
Fig 5-13 Zero Adjustment Page on OMA-2000 Analyzer
Step 3 Press the “F3” button to enter the manual calibration page as shown in Fig 5-14
and start the process of SO
2
calibration. The calibration electro-valve of SO
2
will
be open. Then adjust the flow rate meter for standard gas at bottom left of the
cabinet to be (1.5~2)L/min.
Fig 5-14 Manual Calibration
After the manual range calibration page is displayed, the system automatically opens the

Table of Contents