EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 3-19 Calibration Parameter Menu; Fig 3-20 Calibration Parameter Settings

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 System Operations
25
“Instrument Management Æ Instrument Parameter Æ Calibration Parameter” to enter the
automatic calibration page.
Fig 5-19 Calibration Parameter Menu
Fig 5-20 Calibration Parameter Settings
The meanings of items in the page:
z Calibration density of SO
2
: Set the calibration density of SO
2
z Calibration density of NO: Set the calibration density of NO
z Calibration interval: Set the interval between two automatic calibration;

Table of Contents