EasyManua.ls Logo

FPI CEMS-2000 Series - Fig 3-11 Calibration Parameter Menu

Default Icon
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 System Operations
20
Click the “Prepare” button to open the zero adjustment electro-valve. Then adjust the flow
rate meter for the standard gas at bottom left of the cabinet front panel to control the flow rate at
(1.5~2) L/min and click the “OK” button to remotely adjust the zero point.
5.3.3.3 Auto Zero Adjustment on CEMS-2000 Monitoring Software
You can configure the auto zero adjustment parameters on CEMS-2000 monitoring software
and the system will periodically perform zero adjustment according to the settings. You can click
“Instrument Management Æ Instrument Parameter Æ Calibration Parameter” to enter the
automatic zero adjustment page, as show in Fig 5-11 and Fig 5-12.
Fig 5-11 Calibration Parameter Menu

Table of Contents