EasyManua.ls Logo

G-Tech Pro Competition - Erasing Accelerometer Calibration Data

Default Icon
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144 G-TECH / Pro Competition User Guide, Release 3.1
Erasing Accel. Calibration
Use the following procedure to erase the G-tech’s acceler-
ometer calibration data.
Assuming you have performed two calibrations at
two different temperatures, and the G-tech is held
very still, the Mag reading should always be very
close to +1.000. The closer the G-tech is to
+1.000, the more accurate your measurements will
be.
If the reading is greater than +1.005 or less than
+0.995 you should erase the current calibration
(p. 144) and then re-calibrate the accelerometers
(p. 28).
The X, Y, and Z fields are provided for engineering
support, and generally you will not need to look at
them.
The current temperature inside the G-tech is also
displayed, as this affects the accelerometers. It is
provided primarily for engineering support.
1
Start in Configuration Mode.
2
Select “Calibration” and press .
Note: DO NOT CALIBRATE ACCELEROMETERS or ERASE ACCELEROMETER CALIBERATION 
on G-tech units which were shipped after August 2003. They have already been calibrated in the 
factory. Older G-techs will automatically request calibration. If you are unsure contact tech support.

Table of Contents