REV 5 DIRECTION 2173225-100
10-9
3. There is a large loop of collimator cable immediately after strap point C." See
Illustration 10-11.
ILLUSTRATION 10-11
     
CLAMP POINT A
CLAMP POINT D
STRAP POINT N
STRAP POINT C
4. The collimator cable goes from strap point C" directly into the vertical column
assembly terminal box where it is connected to terminal strip AMX3 A1 TS1. See
Illustration 10-12 and Table 10-1.
ILLUSTRATION 10-12
     
TERMINAL #1
WIRE CABLES
PER CHART
TERMINAL #16
GND STUD
HORIZONTAL ARM BRAKE
AMX3 A2 J1
X-RAY CONTROL
AMX1 A2 TS3
COLLIMATOR
AMX3 A4
X-RAY TUBE
AMX3 A3 TS1
VERTICAL COLUMN
BRAKE CABLE
COLUMN
TERMINAL BOX