This section provides procedures covering the removal and installation of the high
voltage transformer.
!  
  
1. Shut power off by turning the circuit breaker OFF.
2. Remove front cover by carefully pulling cover from four ball stud retainer sockĆ
ets.
3. Disconnect anode and cathode cables from high voltage transformer.
4. Remove two cap screws holding rear cover on base assembly and remove cover.
5. Remove cassette drawer. Refer to Section 5-1.
6. Remove cassette drawer magnetic strip lock by removing three binding head
screws and flat washers.
7. Remove high voltage transformer terminal cover by removing two binding head
screws and washers. See Illustration 9-1.
ILLUSTRATION 9-1
  
  
 
 
 

 
 
 

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode