EasyManuals Logo

GE AMX 4+ Technical Manual

GE AMX 4+
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
 
        
  
6-8
13. Connect one of the Cable #1's end with sense lead to Battery #6 (+), and conĆ
nect other end to Battery #5 (-). Slide boots over each terminal.
14. Connect end of Cable #3 without the sense lead to Battery #6 (-), and slide a
boot over terminal.
15. Push Batteries #5 and #6 to rear of battery compartment.
16. Connect free end of Cable #8 (the cable which comes from battery #4 [-]) to
Battery #5 (+), and slide a boot over terminal.
       
1. Install the battery test terminal strip (see Illustration 6-2) into battery compartĆ
ment as shown in Illustration 6-3.
2. Connect Cable #7 Sense Lead (from Battery #2) to Terminal J2 (RHS [right
hand side]) as shown in Illustration 6-2.
3. Connect Cable #7 Sense Lead (from Battery #4 [+]) to Terminal J4 (RHS).
4. Connect Cable #6 Sense Lead (from Battery #3 [+]) to J3 (RHS).
5. Connect Cable #4 Sense Lead (from Battery #1 [+]) to J1 (RHS).
6. Connect Cable #1 Sense Lead (from Battery #6 [+]) (LHS [left hand side]).
7. Connect Cable #8 Sense Lead (from Battery #5 [+]) to J5 (LHS).
   
1. Install the Insulator #1 between Batteries #5 and #6 and Batteries #7 and #8
with cup-shaped bumps pointing away from you. Route free end of Cable #3
over top of Insulator #1.
2. Position Batteries #7 and #8 in compartment opening stacked one on top of the
other with a separator pad between them.
3. For remaining Cable #1, connect end with sense lead to Battery #8 (+) and conĆ
nect other end to Battery #7 (-). Slide a boot over each terminal.
4. Connect end of Cable #5 without sense lead to Battery #8 (-), and slide a boot
over terminal.
5. Push Batteries #7 and #8 back into position while holding Cable #3 above InsuĆ
lator #1.
6. Connect free end of Cable #3 from Battery #6 [-] vertically to Battery #7 (+)
and slide a boot over terminal.

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals