3-13
  
  
1. Slide ball end of brake cable into slot in cable bracket on end of chain on brake
and cable assembly. See Illustration 3-7.
2. Route anchor end of cable around pulley on end of mounting plate to opposite
end of support bar and insert in slot.
3. Replace anchor nut on cable end and take up slack by turning nut clockwise. Do
not overtighten.
4. Replace cable clip, with brake cable in place, on bar support assembly and secure
with one machine screw.
5. Check that the cable and chain move freely.
6. Loctite the anchor nut.
7. Install horizontal arm brake and cable assembly. Refer to Section 3-8.
  
 
1. Remove horizontal arm brake and cable assembly. Refer to Section 3-7.
2. Disconnect brake electrical cable connectors to remove from connector bracket.
3. Loosen remaining capscrew holding brake to mounting plate. See Illustration
3-7.
4. Slide brake off brake sprocket shaft.
  
 
1. Slide brake onto brake sprocket shaft.
2. Secure brake with single capscrew in end of mounting plate. See Illustration 3-7.
3. Position brake electrical cable connectors in connector bracket and reconnect.
4. 'Install horizontal arm brake and cable assembly. Refer to Section 3-8.

Table of Contents

Other manuals for GE AMX 4+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX 4+ and is the answer not in the manual?

GE AMX 4+ Specifications

General IconGeneral
TypeMobile X-ray System
Maximum kVp125 kVp
CollimatorManual Collimator
BatteryRechargeable, lead-acid
Image Receptor CompatibilityCR and DR systems
Focal Spot Size1.2 mm
mAs Range0.04 - 320 mAs
X-ray TubeRotating Anode

Related product manuals